15
Tecto FI2 84 PI
EN 428001335137
Use and maintenance manual
8. MAINTENANCE
The Staff in charge of the appliance must control and respect the expiry dates for maintenance, given in the table below,
contacting the authorised Technical Assistance when indicated.
ATTENTION
After any maintenance it is mandatory to perform the electric safety tests in compliance with the IEC EN 50106
Standard.
FREQUENCY
Maintenance
Manager
According to Use
and Requirement
Monthly
Every 6 months
Cleaning the external surfaces
USER
Cleaning of accessible internal parts(without the use of tools)
Check integrity of sealing gaskets
Control of power supply cable/plugs and/or electrical sockets
Cleaning the condenser unit lter(If present)
Cleaning the defrosting water collection tray
TECHNICAL ASSISTANCE
Cleaning the water collection tank (inside the tank - if present)
Condenser cleaning
Control of oil level of compressor (Where present)
Venting of air tank (Where present)
Control of pneumatic connections (Where present)
Check the integrity of chiller system piping
Inspect cables and internal power connections
Cleaning of the condensation dryer sponges(Where present)
Replacement of lights / LEDs (Where present)
Replacement of control panel (Electronic control unit - thermostat - etc.)
Replacing the power supply cable, plugs and/or electric sockets
Ordinary Maintenance
Extraordinary Maintenance