Page 44
Typical dose rates (5 milliCurie [185 MBq] source)
with the shutter closed,
in milliREM/hr.,
are as follows:
5 cm 30 cm 100 cm
Left 0.75 0.03 <0.01
Right 0.48 0.025 <0.01
Top 0.25 0.025 <0.01
Bottom 0.55 0.025 <0.01
Front 1.0 0.035 <0.01
Rear 0.05 0.015 <0.01
Typical dose rates (5 milliCuries [185 MBq] source)
with the shutter open,
taking a reading on wood,
in milliREM/hr.,
are as follows:
5 cm 30 cm 100 cm
Left 1.3 0.045 <0.01
Right 1.7 0.06 <0.01
Top 1.9 0.055 0.015
Bottom 0.70 0.04 <0.01
Rear 0.20 0.025 <0.01