6.3 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
- Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
frequency
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power(W)
Distance(m)
Test
385 380 ~ 390 TETRA 400
modulation
b)
1.8 0.3 27
450 430 ~ 470
GMRS 460,
FRS 460
deviation
2. 0.3 28
704 ~ 787
LTE Band 13,
17
modulation
b)
217 Hz
0.2 0.3 9
800 ~ 960
TETRA 800,
iDEN 820,
CMDA 85-, LTE
Pulse
modulation
b)
18 Hz
2 0.3 28
930
1700 ~
1990
CMNA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
Pulse
modulation
b)
217 Hz
2 0.3 28
1970
2450
2400 ~
2570
WLAN, 802.11
b/g/n, RFID
2450, LTE
Pulse
modulation
b)
217 Hz
2 0.3 28
5100 ~
5800
WLAN 820.11
a/n
modulation
b)
0.2 0.3 9
Note. If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is
permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be worst case.