SYLFIRM X User Manual(VO-UM-SFX-01)
Rev. 01 Rev. date: 2020-03-16
Uncontrolled When Printed
All Rights Reserved © 2014. VIOL Co., LTD.
Page 3 of 33
Operating the product may lead to a “spark” which can flare intestinal gas or oxygen-
containing objects (such as cotton and gauze).
14) Patients must not make contact with any grounded metal part or elastic conductor.
Use dry materials to insulate body parts that give out excessive secretion or contact the
skin (e.g. arm, torso).
15) As this product is electrically operated and can be interfered by electromagnetic waves, do
not use other electric devices like cell phone when this product is in use.
16) Since this product is a medical device, instructions are required to properly dispose of it.
Contact the deputy or manufacturer for such instructions. Improper disposing of this device
can lead to legal consequences.
17) To avoid risks associated with electric shock, connect this product to a power supply using
the protective ground connector.
18) Adhere to the following for safely turning off the product.
Reset the device to its default settings when finished operating it.
Do not pull out the plug with one hand using excessive force.
19) Do not set this equipment up at closed place and narrow place, and power plug placed
difficult place to remove.
20) If the equipment fails, the unintended output power may increase or decrease. In this case,
contact the manufacturer.
21) Contact the manufacturer if the product functions abnormally, problematically, or not at all.
Any attempt to repair without the manufacturers guide is not recommended.
22) The conducting wire for patients should be situated in ways that do not make contact with
either the patient or other wires. Any active electrode that is temporally not in use should be
stored in locations that are insulated from the patient.
23) Crusting, Edema may be observed right after treatment, however these symptoms are
temporary and normally expected during and after the treatment.
24) Reuse of the tip is prohibited in any circumstances due to the infection.
25) Re-sterilization is prohibited in any circumstances.
26) Properly shut down, disconnect and secure the hand-piece and foot switch connector before
moving it. Failure to do so may damage the system or cause someone to trip over a loose cable.
27) If any abnormality occurs during use, please contact Manufacturer. Our information is on the
Chapter 8.
28) The safety load weight is as follows.
A. Hand-piece holder: 0.5 kg
B. Hand-piece cable holder: 0.5 kg
C. Cart tray: 10 kg
D. Cart handle: 5 kg