Mantis Compact optical path alignment
Below is the alignment procedure for use in the field. If this procedure fails to align the unit, it is recommended that
the microscope be sent back to the nearest Vision Engineering Service Centre for repair.
Tools required:
3mm, 2.5mm and 2mm hex drivers (ball end recommended).
u Loosen but do not remove the two side screws.
u Loosen the centre screw and use the hex driver as a
"joystick" to re-position the left side optical path until it
horizontally aligns with the right side.
u Use any available straight edge (e.g. a metal rule or strip of tape)
if desired to assist in the visual alignment.