VitaScan LT
User and Service Manual
No annual Inspection and Maintenance is required on your VitaScan LT. If the Monthly
Accuracy Check fails, you should:
• Perform an online integrity and calibration service using a Calibration Test Tool; or
• Perform an o-line verication test using a Verication Test Tool; or
• Contact your local dealer to learn about options for sending in the device for
manufacturer calibration.
Care, Cleaning and Disinfecng
Clean the VitaScan LT with a so cloth soaked in a mild liquid detergent soluon. Rinse with
clean water and carefully dry with a clean so cloth. Dampen a so cloth with 70% ethanol.
Wipe the probe with the dampened so cloth and let the ethanol evaporate.
If the VitaScan probe needs to be disinfected, we recommend CIDEX® OPA Soluon, PDI
Sani-Cloth AF3 Germicidal Disposable Wipes, or other comparable disposable wipe designed
for use on non-porous plasc surfaces. You may also use any glutaraldehyde based hospital
disinfecon soluon or Clorox Healthcare bleach products. Dampen a so cloth and wipe the
instrument thoroughly. To remove all traces of disinfecon soluon, wipe the VitaScan with a
clean so cloth dampened with sterile water or cleaning soluon. Carefully dry the VitaScan
with a clean so cloth before use. Use appropriate hand protecon according to the labeling
on the disinfectant to avoid skin reacons.
Warning:
• Do not subject any part of the VitaScan LT to steam sterilization or ethylene oxide
sterilization.
• Do not immerse the instrument in any cleaning or disinfecting solution.
!
12