20Page of 36
ENG
DT_0006 Issue 15
A Global Leader in Respiratory Solutions
the LED will be orange. The Power Supply unit should be checked regularly and
replaced when necessary.
4.1. Battery Pack
The ALPHA Touch is tted with a rechargeable Battery Pack. This allows the
device to be used without the 12V Power Supply connected. The battery pack
can be re-charged by plugging in the 12 V Power Supply. To fully re-charge
switch off the Vitalograph ALPHA Touch and leave it plugged in over-night.
The battery pack should be replaced after 3 years. Battery replacement should
be carried out only by the manufacturer, the approved importer or by Service
Agents specically approved by Vitalograph.
Note: It is not possible to power the device or charge the batteries from the USB.
4.2. Battery Low Detect
The battery power detect messages are:
Battery Low: Theicon ashes on and off on theMain
Menuscreen.
The Battery Pack is running low. You may continue to use the
device.
As a precaution, plug in the 12V Power Supply to re-charge the
batteries and continue testing.
Battery Discharged: Theicon appears continuously on theMain
Menuscreen. The Battery Pack is out of power. If you press any
key a warning message appears and you will not be allowed to
proceed further.
Plug in the 12V Power Supply to re-charge batteries and continue
testing.
5. Cleaning & Hygiene
5.1. Preventing Cross-Contamination of Subjects
A spirometer is not designed or supplied as a ‘sterile’ device.
Vitalograph intends that a new Bacterial Viral Filter (BVF) be used for every
subject to prevent cross contamination. Using a new BVF provides a signicant
level of protection of for the subject, the device and the user against cross
contamination during spirometry manoeuvres.
The interior of a Vitalograph owhead does not require decontamination where
a new BVF is used for each subject. The outside surfaces of the device and
owhead tube may becleaned with a 70% isopropyl alcohol impregnated cloth
to remove any visible soiling and for low level disinfection.
Where the user suspects that the owhead has become contaminated or where
local risk assessment identies a need for higher level of decontamination, then
it should be cleaned as per the instructions on ‘Cleaning and Hygiene’ on the
Vitalograph website.