EasyManua.ls Logo

Vitalograph ALPHA Touch - Page 31

Vitalograph ALPHA Touch
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
09100 Issue 1
31Page of 36
ENG
ALPHA Touch - Instructions for Use
EN 60601-1-2:2007 - Emissions tests
Emissions test Compliance
Electromagnetic environment -
guidance
RF emissions CISPR 11 Group 1
The Model 6000 ALPHA Touch
uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11 Class B
The Model 6000 ALPHA
Touch is suitable for use in
all establishments, including
domestic establishments and
those connected to the public
mains network (e.g. at home and
doctor's oces in residential
areas)
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/
icker emissions IEC
61000-3-3
Class A
EN 60601-1-2:2007 -Immunity tests
Immunity test Test level Compliance level Reached
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/ output lines
±2 kV for power supply
lines
Surge
IEC 61000-4-5
±1kV differential mode ±2kV
common mode
±1 kV differential mode

Other manuals for Vitalograph ALPHA Touch

Related product manuals