20Page
ENG
DT_0006 Issue 17
A Global Leader in Respiratory Solutions
of 28
EN 60601-1-2:2015 + A1:2021 - Emissions tests
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Model 6600 Compact uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
The Model 6600 Compact is suitable for use
in professional environments only, such as
hospitals, clinics.
This device is not suitable for use in home/
residential environments.
EN 60601-1-2:2015 + A1:2021 -Immunity tests
Immunity test Test level Compliance level Reached
Electrostatic
discharge
(ESD)
EN 61000-4-2
Contact: ± 8 kV Air:
± 2 kV,± 4 kV, ± 8kV
15 kV
Contact: ± 8 kV Air: ± 2 kV,± 4 kV, ±
8kV,± 15 kV
The device may reset/turn off when
exposed to higher levels of ESD. To
recover press the On/Off button.
Proximity
eldsfrom
RF wireless
9to28V/m
385to5785MHz
As per Table 9
EN60601-1-2
9to28V/m
385to5785MHz
AsperTable9EN60601-1-2
Radiated RF
EN 61000-4-3
3 V/m (professional
healthcare)
80MHzto2.7GHz
80%amplitude
modulatedat1kHz
3V/m(professionalhealthcare)
80MHzto2.7GHz
80%amplitudemodulatedat1kHz
Electrical fast
transient/burst
IEC 61000-4-4
±0.5, ±1 & ±2kV
100kHz
±0.5, ±1 & ±2kV 100kHz