09220 Issue 2
21Page of 28
ENG
Vitalograph COMPACT - Instructions for Use
Surge IEC
61000-4-5
±1kV differential
mode Device is
identiedasClassII
unearthed equipment
therefore common
mode testing is
omitted.
±1kV differential mode
Conducted RF
Immunity
EN61000-4-6
3V/m0.15-80MHz
6V/m ISM/amateur
radiosbands
0.15-80MHz
80%AMat1kHz
3V/m0.15-80MHz
6V/mISM/amateurradiosbands
0.15-80MHz
80%AMat1kHz
Voltage
dips, short
interruptions
on power
supply input
lines IEC
61000-4-11
100%drop,0.5cycles,
0o, 45o, 90o, 135o,
180o, 225o, 270o,
315o
100%dip,1cycle
30%dips,25/30
cycles
100%drop,0.5cycles,0o,45o,90o,
135o,180o,225o,270o,315o
100%dip,1cycle
30%dips,25/30cycles
Proximity
magnetic
Fields
EN61000-4-39
8A/m30kHz
65A/m134.2kHz
(2.1kHzPM)
7.5A/m13.56MHz
(50kHzPM)
8A/m30kHz
65A/m134.2kHz
(2.1kHzPM)
7.5A/m13.56MHz
(50kHzPM)
Power
frequency
(230V,
50Hz and
120V, 60Hz)
magneticeld
30A/m 30A/m
Medical Devices may be affected by mobile RF communications equipment
including cellular telephones and other personal or household devices not
intended for medical facilities. It is recommended that all equipment used
near the Vitalograph product comply with the medical electromagnetic
compatibility standard and to check before use that no interference is
evident or possible. If interference is suspected or possible, switching off the
offending device is the normal solution, as is required in aircraft and medical
facilities.