B Measuring tolerances
Indication of the measuring tolerances Cx in ± (% of the reading + number of
digits = digit error)
Type Measuring Precision Dissolution Calibration of
Operation Range (SET mode)
Capacity 20 nF ±(1.0%+5dgts) 1 pF open
with 120 Hz 200 nF ±(0.7%+5dgts) 10 pF open
test frequency 2 µF ±(0.7%+3dgts) 100 pF ---
20 µF ±(0.7%+3dgts) 1 nF ---
200 µF ±(0.7%+3dgts) 10 nF ---
2000 µF ±(1.0%+5dgts) 0.1 µF short-circuit
(short)
20 mF ±(5.0%+5dgts) 1 µF short-circuit
Capacity 2000 pF ±(1.0%+5dgts) 0.1 pF open
with 1 kHz 20 nF ±(0.7%+5dgts) 1 pF open
test frequency 200 nF ±(0.7%+5dgts) 10 pF ---
2000 nF ±(0.7%+3dgts) 100 pF ---
20 µF ±(0.7%+3dgts) 1 nF ---
200 µF ±(1.0%+3dgts) 10 nF short-circuit
2000 µF ±(5,0%+5dgts) 1 µF short-circuit
pF = pico-Farad = 10 exp. -12; nF = nano-Farad = 10 exp. -9; mF = micro-
Farad = 10 exp.-6
=========================================================
Inductivity 20 mH ±(2.0%+5dgts) 1 µH short-circuit
with 120 Hz 200 mH ±(1.0%+5dgts) 10 µH short-circuit
test frequency 2000 mH ±(0.7%+5dgts) 100 µH ---
20 H (< 0.5H) ±(0.7%+5dgts) 1 mH ---
200 H (> 0.5H) ±(0.7%+5dgts) 10 mH ---
2000 H ±(1.0%+5dgts) 100 mH open
20000 H not specified 1 H ---
Inductivity 2000 µH ±(2.0%+5dgts) 0.1 µH short
with 1 kHz 20 mH ±(1,2%+5dgts) 1 µH short
test frequency 200 mH ±(0.7%+5dgts) 10 µH ---
42
H Display symbol for the hold-function
SET Display symbol for the setting mode
REL Display symbol for measuring relative to a reference value
TOL Display symbol for tolerance value measuring
PAL Display symbol for the parallel mode (parallel connection)
SER Display symbol for series mode (connection in series)
D Display symbol for the dissipation factor D = 1/Q
Q Display symbol for the quality
R Display symbol for the parallel or series loss resistor
% Display symbol for the tolerance in [%]
1 kHz Display symbol for the measuring frequency 1 Kilohertz
120 Hz Display symbol for the measuring frequency 120 Hertz
L C R Display symbols for measuring the inductivity "L" (inductor),
the capacity "C" (capacitor) and the ohmic resistance "R" (resistor)
^ Display symbol for increasing the set value, or for displaying the
upper tolerance
Display symbol for decreasing the set value, or for displaying the
lower tolerance
"" Low battery display symbol
M K Ω Display symbols for the measuring units for resistor measuring
µmH Display symbols for the measuring units for inductor
measuring
mµnpF Display symbols for the measuring units for capacitor measuring
Table of Contents
Introduction ...................................................................................24
Prescribed Use ..............................................................................25
Setting Elements (fold-out page)...................................................25
Table of Contents ..........................................................................27
Safety Instructions.........................................................................28
Presentation...................................................................................29
Description of the Function and the Buttons ................................30
Disposal.........................................................................................38
Trouble-Shooting...........................................................................38
Maintenance and Calibration.........................................................39
Technical Data and Measuring Tolerances ...................................41