Safety Instructions
Any claims for guarantee will become invalid in the event of damage that
results from the non-observance of the operating manual. We do not accept
responsibility for such damage nor for any indirect losses.
We do not accept responsibility for damage to property or for personal
injuries caused by improper use or non-observance of safety instructions.
Guarantees will not be accepted in any such case.
- The measuring device has been constructed and tested according to the
standard EN 61010-1, Protective Measures for electronic Measuring
Instruments, and has left the factory in a perfectly safe condition.
- We kindly request the user to respect this operating manual in order to pre-
serve this condition and to ensure safe operation. Keep measuring devices
out of the reach of children!
- Observe the safety regulations laid down by the professional trade associ-
ation for electrical equipment and installations in industrial and commercial
facilities.
- Measuring devices and accessories used at schools, training facilities, do-
it-yourself and hobby workshops must be supervised by the responsible
qualified personnel.
- Be especially careful when handling voltages over 25 V AC or over 35 V
DC. Even with these voltages, you can suffer a dangerous electric shock
when you get in contact with electrical leads.
- Every time before you change the measuring range, the measuring pivots
must be removed from the measured object.
- Check the measuring instrument and the measuring leads for damages
before every use!
- Before every measurement the capacitors that are to be measured must be
completely discharged.
- Any measuring of live components, circuit elements etc. is prohibited. Dan-
ger to life! under contact with voltages over 25 V ACrms or 35 V DC.
- Do not operate the measuring instrument in rooms or under adverse envi-
ronmental conditions, where there are or could be inflammable gases,
vapours or dust.
28
Technical Data and Measuring Tolerances
A Technical Data
Display ......................................: two 4
1
/
2
digit displays up to 19999 with sym-
bol display and measuring units
Maximum measuring rate .........: 1 measurement per second
Overload display .......................: ".OL” for Overload is displayed
Automatic Power Off APO ........: APO for Automatic Power Off, can be de-
activated, and does not function when the
interface is operated, in the MIN-MAX func-
tion and when supplied by an external
adapter.
The measuring instrument will switch off
automatically, when it is not used for about
10 minutes. To switch in on
again, press the power button
Operating temperature..............: 0°C to +50°C
Relative humidity in air..............: 0 to 80%, not condensating
Storage temperature
(battery removed)......................: - 20°C to + 60°C, with a rel. hum. of 0 to
80%
Temperature for guaranteed
precision ...................................: +23°C ±5 K
Low battery signal.....................: " " from approx. under 7 V battery
voltage
Battery type ..............................: NEDA 1604 9V or 6F22 9V (alcaline)
or MN 1604 (=PP3)
Current consumption................: approx. 10 mA to approx. 30 mA
External power supply ..............: nom. 12 V DC, min. 50 mA, polarity "+” inside
Fuse, defect signal....................: 100 mA, fast 250 V, usual denomination
F0.1/250V (5 x 20 mm).
When the fuse is defective, you will hear an
uninterrupted signal sound when you switch
on the instrument, and "FUSE" will appear in
the display.
Weight.......................................: 365 g (including battery and rubber holster)
Dimensions (L x W x H).............: 192 x 91 x 52.5 mm (without rubber holster)
-
+
41