www.vtiinstruments.com
EX1200 Series Maintenance 145
LAN/GPIB Gateway
PC
LAN Port
Ethernet
Cable
GPIB
Cable
GPIB
Cable
Daisy
Chain
GPIB
Cable
Daisy
Chain
Interconnect Cable
Function Generator
8.5 Digit DMM
Amplifier
Calibrator
LAN Port
GPIB Port
GPIB Port
GPIB Port
GPIB Port
Ethernet
Switch
Ethernet
Cable
LAN Port
FIGURE 9-4: INSTRUMENT CONNECTIVITY FOR EX1200 DMM CALIBRATION
Once the standards have been connected, the following connections should be made to the
EX1200:
8) Connect the EX1200 LAN port to the Ethernet switch using an Ethernet cable.
9) Connect the function generator, external DMM, and calibrator to the DMM as shown in
Figure 9-5 using the EX1200 DMM calibration/verification cable (VTI P/N: 52-0807-000).
External DMM
Input I
Input HI
Input LO
Sense HI
Sense LO
Function Generator
Output HI
Output LO
Sync HI
Sync LO
Calibrator
Output I
Output HI
Output LO
Sense HI
Sense LO
EX1200 DMM
Input I
Input HI
Input LO
Sense HI
Sense LO
FIGURE 9-4: EX1200 CONNECTIVITY TO REFERENCE STANDARDS
NOTE Be certain that the calibrator is in standby mode (i.e. not outputting) before the cables are
connected to eliminate the possibility of having the function generator and calibrator generating
outputs at the same time while shorted together.
CONNECTING TO THE EX1200
There are three ways to connect to the EX1200: using the IP address of the EX1200, using
Bonjour for auto-discovery, or using the Agilent Connection Expert (ACE). For more information
on obtaining the EX1200’s IP address and using this as the connection method, refer to the
Network Configuration discussion in Section 2. To connect to the EX1200 via Bonjour, refer to
Section 7: Web Page Operation.