148 EX1200 Series Maintenance
Calibration Procedure
1) Connect to the embedded web page. See Connecting to the EX1200 for assistance.
2) Click Soft Front Panel from the Index page to navigate to the Soft Front Panel page. Select
the ex1200-dmm DMM from the list of available plug-in modules (the slot designation will
vary depending on the mainframe being used).
FIGURE 9-8: INSTRUMENT SFP SELECTION PAGE
3) At the top of the DMM’s soft front panel, click the Calibration button.
FIGURE 9-9: CALIBRATION TAB
4) From the Calibration Functions section of the Calibration page, select Calibration from
the drop down menu.
FIGURE 9-10: CALIBRATION FUNCTION SELECTION
Select the DMM
Calibration Tab