12 VWR MEGA STAR 3.0 / 3.0R ver b, rel 25.06.2020
Product Specifications
Characteristics of the VWR MEGA STAR 3.0 / 3.0R
Several rotors with a wide range of tubes can be used.
The set speed is reached within seconds. The maintenance-free induction motor ensures quiet and
low-vibration operation even at high speeds and guarantees a very long lifetime.
The user-friendly control panel makes it easy to pre-set the speed, RCF value, run time and temperature.
You can choose between the display of speed and RCF or the entry mode.
These settings can be changed even while the centrifuge is running.
With the help of the PULSE key , you can also centrifuge a sample for just a few seconds, if called for.
The VWR MEGA STAR 3.0 / 3.0R is equipped with various safety features:
• The housing is made of impact-resistant plastic and the interior of armour steel.
• The lid is equipped with a view port and a lock.
• The lid of the centrifuge can only be opened while the centrifuge is switched on and the rotor has
come to a complete stop. The centrifuge cannot be started until the lid has been closed properly.
• Lid emergency release: For emergencies only, e.g. during power failures.
Contains fluorinated greenhouse gases in a hermetically sealed system.
Mains supply
The following table contains an overview of the electrical connection data for the VWR MEGA STAR
3.0 / 3.0R. This data is to be taken into consideration when selecting the mains connection socket.
Cat. Mains voltage Frequency Rated
current
Power
consumption
Equipment fuse Building
fuse
521-1751 230 V ±10 % 50 / 60 Hz 11 A 1700 W 4 A, 2-pole circuit
breaker
16 AT
521-1752 230 V ±10 % 50 / 60 Hz 10.5 A 1950 W 4 A, 2-pole circuit
breaker
16 AT
Refrigerant
Order No. Centrifuge Refrigerant Quantity Pressure GWP CO
2
e
521-1752 VWR MEGA STAR 3.0R R-134a 0.43 kg 21 bar 1430 0.61 t