13 VWR MEGA STAR 3.0 / 3.0R ver b, rel 25.06.2020
Technical data
The technical data of the VWR MEGA STAR 3.0 / 3.0R is listed in the following table.
Feature MEGA STAR 3.0 MEGA STAR 3.0R
Environmental conditions - Use in interior spaces.
- Altitudes of up to 2,000 m above sea level.
- max. relative humidity 80 % up to 31 °C; decreasing linearly up to
50 % relative humidity at 40 °C.
Permissible ambient temperature
during operation
+2 °C to +35 °C +2 °C to +35 °C
Overvoltage category II II
Pollution degree 2 2
Heat dissipation 230 V 230 V
5800 BTU/h 6653 BTU/h
IP 20 20
Run time unlimited unlimited
Max speed n
max
15,200 rpm (depending on the
rotor)
15,200 rpm (depending on the
rotor)
Min speed n
min
300 rpm 300 rpm
Maximum RCF value at n
max
25,314 x g 25,314 x g
Maximum kinetic energy 62.5 kJ 51.7 kJ
Noise level at maximum speed < 64 dB (A) < 64 dB (A)
Temperature setting range -10 °C to +40 °C
Dimensions
Height
Width
Depth
Table top height
360 mm
550 mm
670 mm
310 mm
360 mm
745 mm
670 mm
310 mm
Weight with empty rotor 86 kg 116 kg