34 VWR MEGA STAR 3.0 / 3.0R ver b, rel 25.06.2020
7. Be sure the decontamination solution can drain off the rotor.
8. Rinse the rotor and accessories thoroughly with water.
9. Dispose of the decontamination solution according to the applicable guidelines.
10. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50 °C.
After cleaning, treat the entire surface of aluminium parts with corrosion protection oil (521-1769).
Also treat the cavities with oil.
Treat the bolt of the swing out rotor with bolt grease (521-1770).
Autoclaving
1. Before autoclaving clean rotor and accessories as described above.
2. Place the rotor on a flat surface.
Rotors and adapter can be autoclaved at 121 °C.
The maximum permissible autoclave cycle is 20 minutes at 121 °C.
Note: No chemical additives are permitted in the steam.
CAUTION Never exceed the permitted temperature and duration when autoclaving. If the rotor
shows signs of corrosion or wear, it must be replaced.
Service of VWR
VWR recommends having the centrifuge and accessories serviced once a year by an authorized service
technician. The service technicians check the following:
the electrical equipment;
the suitability of the set-up site;
the lid lock and the safety system;
the rotor;
the fixation of the rotor and the drive shaft.
VWR offers inspection and service contracts for this work. Any necessary repairs are performed for free
during the warranty period and afterwards for a charge.
This is only valid if the centrifuge has only been maintained by a VWR service technician.