Before use
Before setting up
2-2 VWR MICRO STAR 17 / 17R
Before setting up
1. Check the centrifuge and the packaging for any shipping damage.
Inform the shipping company and VWR immediately if any damage is discovered.
2. Remove the packaging.
3. Remove the rotor transport protection. The cardboard is suppose to prevent damage due to
impact.
4. Check the order for completeness (see “Scope of supplyon page iii).
If the order is incomplete, please contact VWR.
Transporting the centrifuge
Due to its weight (see “Technical dataon page 1-2), the centrifuge should be carried by several
people.
Always lift the centrifuge at both sides on the bottom plate.
Figure 2-1.
Lifting the centrifuge at both sides.
The centrifuge can be damaged by impacts.
Location
The centrifuge should only be operated indoors.
The set-up location must fullfil the following requirements:
A safety zone of at least 30 cm must be maintained around the centrifuge.
People and hazardous substances must be kept out of the safety zone while centrifuging.
WARNING Always lift the centrifuge on both sides. Never lift the centrifuge by its front
or the back panel.
Transport the centrifuge if possible in its packaging and with the rotor transport
protection.
CAUTION Refrigerated centrifuges must be left idle at the new location for about an hour
in order for the coolant to settle in the compressor. Do not start the centrifuge in the time.