31
Apnea Backup in APRV/BiPhasic and CPAP/PSV
When these modes are selected, you must:
1. Set the primary and advanced settings for CPAP/PSV or APRV/BiPhasic.
2. Select the breath type for Apnea Backup mode (Volume or Pressure in adult and
pediatric patients or Volume, Pressure or TCPL in neonatal patients) by pressing the
Apnea Settings button.
3. Set the primary controls appearing at the bottom of the touch screen, for the selected
apnea breath type before pressing the MODE ACCEPT button.
Note: The controls for apnea backup ventilation are not visible once the MODE
ACCEPT button has been pressed. Only the controls that are active and required
for the selected mode display on the main screen once the MODE ACCEPT button
is pressed.