  • E
    Erin MartinezAug 15, 2025
    How to fix a frozen display on Vyaire MicroLab Medical Equipment?
    • J
      Jennifer KelleyAug 15, 2025
      If the display 'freezes' and your Vyaire Medical Equipment unit doesn't respond to key presses, it might be because multiple icons were selected or accidentally pressed. To resolve this, hold the on/off button down for 10 seconds until the unit switches off, and then turn it on again.
  • S
    Suzanne FoxAug 18, 2025
    Why my Vyaire MicroLab Medical Equipment battery does not hold charge?
    • L
      Lisa CastilloAug 18, 2025
      If the Vyaire Medical Equipment battery doesn't hold a charge, the battery might be exhausted, requiring the unit to be returned for servicing. Alternatively, the mains charger could be faulty, in which case you should replace it.
  • P
    Patty WhiteAug 21, 2025
    What to do if stylus does not register icons on Vyaire MicroLab?
    • P
      Paul SmithAug 21, 2025
      If the stylus doesn't register icons on the Vyaire Medical Equipment display, the touch screen might need calibration. Select the calibration check icon, choose touch screen, and follow the instructions provided.
  • J
    Jonathan WagnerAug 25, 2025
    Why are blows inverted on the display of my Vyaire MicroLab?
    • P
      plopezAug 26, 2025
      If your Vyaire Medical Equipment shows blows inverted on the display, it could be due to a broken head assembly or cable. Consider replacing the head assembly or returning the unit for servicing.
  • W
    waresaraAug 27, 2025
    What to do if Vyaire MicroLab Medical Equipment does not register a blow?
    • J
      John PadillaAug 27, 2025
      If the Vyaire Medical Equipment is not registering a blow, it might be due to a broken head assembly or cable. You may need to replace the head assembly or return the unit for servicing.
  • M
    Michael WhiteAug 30, 2025
    Why blows tracing ends abruptly on Vyaire Medical Equipment?
    • L
      Lonnie JohnsonAug 31, 2025
      If the blows tracing ends abruptly while the patient is still exhaling on your Vyaire Medical Equipment, the turbine might be sticking. Try cleaning the turbine in warm soapy water or a sterilizing solution. If the problem persists, you may need a replacement turbine.
  • J
    Jose CookSep 2, 2025
    What to do if there is no display present on Vyaire MicroLab Medical Equipment?
    • W
      William GrossSep 2, 2025
      If there's no display present on your Vyaire Medical Equipment, the charger might not be connected, or the battery could be exhausted. Connect the charger to the mains and let the unit fully charge, or return the unit for servicing.
  • M
    matthewstephensSep 5, 2025
    How to fix paper not printing issue on Vyaire MicroLab Medical Equipment?
    • R
      Robert GreeneSep 5, 2025
      If paper is not printing on your Vyaire Medical Equipment, ensure the paper is correctly housed in the printer compartment. Also, confirm you are using Vyaire-recommended thermal paper.
  • W
    wandaandersonSep 8, 2025
    What to do if icons missing from the display of Vyaire Medical Equipment?
    • M
      Maria MccartySep 8, 2025
      If icons are missing from the display of your Vyaire Medical Equipment, it's possible that the icon has been deselected. Hold the stylus onto the blue area of the display; a list will appear, allowing you to ensure the required icon is selected.
  • S
    Stephanie KellySep 12, 2025
    Why calibration procedure failed on Vyaire MicroLab Medical Equipment?
    • B
      Barbara YorkSep 12, 2025
      If the calibration procedure fails or cannot be completed on your Vyaire Medical Equipment, it could be due to several reasons. The turbine may be faulty, so repeat the calibration procedure; if the problem persists, replace the turbine or return the unit for servicing. Ensure the syringe is fitted to the turbine using the adapter supplied. Also, ensure you are using a Vyaire syringe. The syringe should be emptied and filled with one smooth stroke, avoid pushing down on the shaft or banging at the end of each maneuver.

Summary

I. Introduction - English

II. Introduction - English - USA

Contraindications

Lists contraindications and conditions for suboptimal spirometry testing.

Warnings and Cautions

Defines terms like CAUTION, WARNING, and NOTE for safe device operation.

Getting Started

Guides users through patient selection, data entry, and test execution workflows.

Calibration Check (Verification)

Details the procedure for performing a calibration verification using a 3L syringe.

PC connection using SPCS

Describes connecting the MicroLab to a PC via USB or serial port using SPCS.

Disinfection with Perasafe

Offers step-by-step instructions for disinfecting the spirometer using Perasafe solution.

Trouble Shooting Information

Provides a troubleshooting guide for common operational problems and their solutions.

Vyaire MicroLab Specifications

General IconGeneral
ModelMicroLab
CategoryMedical Equipment
ManufacturerVyaire
DisplayLCD
ConnectivityUSB

Related product manuals