Safety Designation per IEC 60601-1
Type of protection against elec-
trical shock
Internally powered
Equipment and Class I
Degree of protection against
electrical shock
Type B applied part
Power Equipment Battery type: NiMH battery
pack, 8.4V, 1100mAh
Degree of Electrical connection
between equipment and Patient
Equipment designed as
non-electrical connection to
the patient.
Degree of mobility Transportable
Mode of operation Continuous
Classications according to IEC 60601-1
MicroLab Applied part, type B
Volume Transducer Applied part, type B
Option SPO2 Finger Probe Applied part, type BF
(Nonin IPOD
®
Integrated Oximetry Device)
WARNING: No modication of this equipment is allowed.
Note: When you connect other equipment to the unit, always
make sure that the whole combination complies with the
international safety standard IEC 60601-1 for medical electrical
systems. During measurements, connect the MicroLab only to
printers and computers that comply with IEC 60601-1 / ANSI/AAMI
ES60601-1:2005 / CAN/CSA-C22.2 No. 60601-1:14.
WARNING: The user must not touch any voltage-carrying parts
and the patient at the same time.
During database upload the MicroLab may be connected to a
computer that complies with EN 60950 – ‘Information technology
equipment - Safety - Part 1: General requirements’.
87