Electromagnetic Compatibility (EMC) to IEC
60601-1-2
WARNING: Use of portable phones or other radio frequency
(RF) emitting equipment near the system (< 30cm) may cause
unexpected or adverse operation
The MicroLab has been tested to IEC 60601-1-2:2014, regarding it’s
ability to operate in an environment containing other electrical/electronic
equipment (including other medical devices).
The purpose of this testing is to ensure that the MicroLab is not likely to
adversely affect the normal operation of other such equipment and that
other such equipment is not likely to adversely affect the normal operation
of the MicroLab.
Despite the testing of the MicroLab that has been undertaken, normal
operation of the MicroLab can be affected by other electrical/electronic
equipment and portable and mobile RF communications equipment.
Keep a distance of about 2 meters from possible error sources when using
the device.
As the MicroLab is medical equipment, special precautions are needed
regarding EMC (electromagnetic compatibility).
It is important that the MicroLab is congured and installed/put into
service, in accordance with the instructions/guidance provided herein and
is used only in the conguration as supplied.
Changes or modications to the MicroLab may result in increased emissions
or decreased immunity of the MicroLab in relation to EMC performance.
The MicroLab should be used only with the accessories (USB cables, mains
adapter and turbine transducer) supplied (which are referenced in the
accessories section of this manual). None of the MicroLab cables should be
extended in length by the user.
If any cables are extended by the user or non approved accessories are
used, this may result in an increased level of emissions or decreased level
of immunity, in relation to the MicroLab’s EMC. None of the MicroLabs
accessories should be used with other devices, as this may result in an
increased level of emissions or decreased level of immunity, in relation to
the other devices’s EMC.
Part Number Description
36-CAB1048 USB Lead (MicroLab to PC)
36-CAB1054 Power line EU to mains adapter
List of EMC important cables. Do not exchange by other types.
42