WARNING: The MicroLab should not be used adjacent to or stacked
with other equipment. If adjacent or stacked use with other equipment is
necessary, the MicroLab and the other equipment should be observed /
monitored, to verify normal operation in the conguration in which it will
be used.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The MicroLab is intended for use in the electromagnetic environment specied below.
The customer or the user of the MicroLab should assure that it is used in such an
environment.
Emission Test Compliance Electromagnetic Enviroment -Guidance
RF emissions
CISPR 11
Group 1 The MicroLab uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Group B The MicroLab is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes
Harmonic emissions
IEC61000-3-2
Class A
Voltage uctuations/
icker emissions
IEC61000-3-3
Complies
The MicroLab has an essential performance – the product should continue
to operate correctly. In the unlikely event of a Fast Transient / ESD event
occuring, the device should be reset and located away from the source of
interference.
36-CAB1121 Power line Brazil to mains adapter
36-CAB1057 Power line UK to mains adapter
36-CAB1051 Power line AUS/NZ to mains adapter
36-CAB1060 Power line US to mains adapter
43