51kHz This is the operating frequency of the transducer. In this example it
is 51 kilohertz. This represents a vibration of 51,000 times a second.
20
o
C This is the current temperature of the solution in the tank. It is
rounded to the nearest degree centigrade. A reading of 20
o
C means
the temperature is between 19.6
o
c and 20.5
o
c (inclusive).
02:26 This means the unit has another 2 minutes and 26 seconds to run
before the current cycle is complete. This will count down until 00:00
when the cycle will be complete and the unit will stop.
When the timer reaches 00:00 the cycle will stop and one of the two
following screens will be displayed:
XXXXXXXXX represents the unique validated cycle number.
If there was a problem with the cycle the following screen is displayed:
Again XXXXXXXXX represents the unique cycle number.
The cycle could have failed for many reasons. A complete list of possible
reasons can be found on page 22 of this handbook.
After a failed cycle the instruments should be removed, rinsed in clean cold
water and the cycle repeated immediately.
Press the key to return to the home screen.