Page
maintenance…
THE UNIT CONTAINS NO USER SERVICEABLE PARTS.
NO ATTEMPT SHOULD BE MADE TO ENTER THE UNIT
AS THIS MAY INVALIDATE ANY WARRANTY AND/OR
SERVICE CONTRACT WARRANTY.
At the end of each daily session the unit should be emptied and the tank
rinsed thoroughly with clean water. The lid and basket should also be
rinsed with clean water. Both the lid and the tank should then be sprayed
with a non-corrosive bactericidal solution and wiped with disposable paper
towel or just wiped with a non-corrosive bactericidal wipe. The unit should
then be dried thoroughly with disposable paper towel. It is advisable to
leave the lid OFF to allow the unit to dry thoroughly.
The outer case can be wiped with a clean damp microfibre cloth and then
again must be thoroughly dried with disposable paper.
Once a week the outer case of the unit should be sprayed with a non-
corrosive bactericidal solution and wiped with a clean damp microfibre
cloth, and then thoroughly dried with disposable paper towel. The case
should then be wiped with a disposable paper towel and a small amount
of mineral oil. It should be wiped in the direction of the ‘grain’ of the metal
(horizontally). Care should be taken not to allow large amounts of oil to
build up around the edges of the front or rear labels as this may lead to
degradation of the adhesive.
Current evidence suggests that Alcohol is shown to bind blood and
proteins to stainless steel therefore the use of Alcohol wipes is NOT
advised.