Service Manual
OLP-101-15
Settings
OLS:
Lambda
=
1550 nm, level
=
0 dBm, FMOD
=
CW
OLA-100:
Lambda coupled to OLS, see below for attenuation
OLP-110:
ABS, STD, Lambda coupled to OLS
FMOD
=
CW,
BW
=
2
HZ,
AUTORANGE ON,
DlSPL
=
x.xx
DUT Lambda
=
1550 nm
Measurement
-
Carry out the reference measurement with the OLP-110 at 1550 nm and -3 dBm.
Set the OLS and the OLA so that the OLP-110 indicates -3.00 dBm.
-
Carry out measurement with the DUT.
-
Repeat the measurements at -30 dBm and -50 dBm. For function check limits see table
below.
Level
-3 dBm
-30 dBm
8.5
Measurement Error at 850 nm
OLP-15:
The check is carried out at 850 nm
+
2
nm.
Specified limits
f
0.70 dB
f
0.70 dB
Level
-3 dBm
Note: The reference measurement using the OLP-110 is only valid if the wavelengths of the
OLS and the OLP-110 are coupled (automatic correction of the wavelength dependency for the
reference measurement).
Uncertainty of
measurement
f
0.20 dB
f
0.20 dB
Specified limits
f
0.22 dB
Function check
limits
f
0.90 dB
f
0.90 dB
Uncertainty of
measurement
f
0.20 dB
Function check
limits
f
0.42 dB