Page #18 background imageLoading...
Page #18 background image
Service
Manual
OLP-101-15
Settings
OLS: Lambda
=
1310 nm, level
=
0 dBm,
FMOD
=
CW
OLA-100:
Lambda coupled to OLS, see below for attenuation
DUT: Lambda
=
131 0 nm
Measurement
-
Set the
OLA
so that the DUT indicates -20 dBm (approx.).
-
Switch on the OLS frequency modulation with FMOD
=
270 Hz, 1 kHz and
2
kHz one after
the other.
The DUT should display the frequencies.
-
Repeat the check at -47 dBm CW.
Note: With modulated light the displayed level is
3
dB
lower than in CW.