18-2 Model 9100 — Operator's Guide: Error Messages
Type C Overload Errors
Overload errors occur when the Model 9100 detects
over-current, over-voltage, over-temperature or over-
compliance conditions in its analog output circuits.
Action required:
Remove the source of the overload, and in the case
of over-temperature conditions wait for the Model
9100 to cool down. If the overload condition persists,
contact your local Wavetek Service Center.
Type D Calibration Errors
Although most of these errors occur during calibration
of the Model 9100, they may also occur when the
Model 9100 interrogates its calibration stores during
normal operation.
Action required:
Error messages indicating a password failure or
incorrect cal switch position can be overcome by
rectifying the condition before proceeding. If the
error message indicates a corrupt calibration store,
recalibrate the Model 9100. If the failure message
persists, contact your local Wavetek Service Center.
Type E Cold Junction Compensation
Errors
These errors occur when the thermistor which senses
the temperature of the Model 9100's terminals
indicates an obvious temperature error.
Action required:
Contact your local Wavetek Service Center.
Type F Configuration Errors
These errors only occur when you are configuring
the Model 9100 using the CONFIG screen menus.
Action required:
Correct the configuration data which has caused the
error and proceed.
Error Messages