9 Alarms and error messages
9.1.2 Muting the alarm
If an alarm sounds, you can mute the audible alarm for 2 minutes.
Requirement An alarm has been triggered.
1. Press the mute symbol .
Result The alarm is muted for 2 minutes. The symbol turns orange. After
2 minutes, the audible alarm sounds again.
.
9.1.3 Pausing the alarm
If an alarm sounds, you can pause the alarm for 2 minutes to
operate the device normally in the meantime.
Requirement The Apnea, Low minute volume, or Low tidal volume alarm
has been triggered.
1. Press the PAUSE field.
Result The alarm is paused for 2 minutes. The symbol appears in the
status line. After 2 minutes, the audible alarm sounds again.
.
Low minute volume! Please
check the ventilation settings
and the course of the
respiration hose.
The respiratory volume output by
the device is lower than the
target value.
Check that the
respiration
hose is neither
blocked nor kinked. Reposition the mask
and breathe through it. If the alarm
continues to show: Have the settings
checked by the attending physician.
Alarm message Cause Remedy
Pressure build-up not possible!
Please connect the mask and
hose.
No respiration hose and/or mask
connected.
Connect the mask and respiration hose
correctly (see "4.2.2 Connecting up the
respiration hose", page 32).
If your physician has activated this function, you can also
deactivate the Severe leakage alarm permanently (see "6.4
Setting device parameters", page 65).
If your physician has activated this function, you can also
deactivate the Severe leakage alarm permanently (see "6.3
Setting time parameters", page 64).