WM 67841c 03/2015
WM 100 TD EN 85
12 Appendix
12 Appendix
12.1 Technical data
12.1.1 Technical data on therapy device
Specification Therapy device
Product class according to 93/42/EEC IIa
Dimensions W x H x D in cm 17 x 13.5 x 18
Weight 1.4 kg
Temperature range
- Operation
- Storage
+5°C to +40°C
-25°C to +70°C
Permissible humidity during operation and
storage
Rel. humidity 15% to 93%, non-condensing
Air pressure range
700 hPa to 1060 hPa, corresponds to a height of 3000 m
above sea level
Connection diameter of respiration hose in mm 19.5 (to fit standard cone)
Electrical power Max. 40 VA
System interface
12 V DC
Max. 10 VA
Current consumption during operation (Therapy)
230 V
115 V
during standby mode (Standby)
230 V
115 V
0.11 A
0.22 A
0.036 A
0.019 A
Classification acc. to DIN EN 60601-1-11:
Protection class against elec. shock
Degree of protection against elec. shock
Protection against harmful ingress of water and
foreign bodies
Protection class II
Type BF
IP21
Classification as per DIN EN 60601-1:
Operating mode
Continuous operation
Applied part Respiratory mask