Physician menu
SOMNObalance (e) / SOMNOsoft 2 (e) EN 7
WM 67441 07/10
3 Physician menu
3.1 Navigate in the Physician menu
1. Connect the device to the electricity supply.
The device is on standby.
2. Press the softstart key and keep it depressed.
The first parameter in the patient menu appears first (SoFT).
3. To reach the Physician menu, continue to keep the sofstart key
depressed and then also press the humidifier key .
After about 4 seconds, the first parameter starts flashing and
can be set.
4. Press the on/off key (
+) to increase the value.
5. Press the softstart key (-) to reduce the value.
Tip If you keep the relevant key depressed, the speed of adjustment
accelerates.
6. To scroll through the menu and switch to the next parameter,
press the humidifier key .
The therapy device adopts the setting.
7. To exit the Physician menu, press the info key.
Alternatively,
wait 15 seconds. If you do not press any key for 15 seconds, the
device automatically exits the Physician menu.