Physician menu
8 EN SOMNObalance (e) / SOMNOsoft 2 (e)
WM 67441 07/10
3.2 Parameters in the Physician menu
Parameter Display
on
device
Value range
Therapy mode
APAP
CPAP
CPAP, APAP (SOMNObalance (e) only)
If CPAP: CPAP
pressure
P
SOMNObalance / SOMNOsoft 2:
4 hPa - 18 hPa
SOMNObalance e / SOMNOsoft 2e :
4 hPa - 20 hPa
If APAP: lower
pressure limit
P
min
4 hPa-P
max
If APAP: upper
pressure limit
P
max
SOMNObalance / SOMNOsoft 2:
P
min
-18 hPa
SOMNObalance e / SOMNOsoft 2e :
P
min
-20 hPa
Rate of pressure
rise (APAP)
0. hPa - 0.4 hPa/s
softPAP level
Soft
0 (none/off)
1 (slight drop)
2 (normal drop)
softPAP lock
SLOC
0 (lock disabled, patient can change
softPAP setting)
1 (lock enabled, patient cannot
change softPAP setting)