Troubleshooting
Lead-quality problems
“Artifact” message on the screen
Artifact is signal distortion that makes it difficult to accurately discern the waveform
morphology.
Causes
The patient was moving.
The patient was shivering.
There is electrical interference.
Actions
See actions for wandering baseline, muscle tremor, and AC interference.
Wandering baseline
Wandering baseline is an upward and downward fluctuation of the waveforms.
Causes
Electrodes are dirty, corroded, loose, or positioned on bony areas.
The electrode gel is insufficient or dried.
The patient has oily skin or used body lotions.
Rising and falling of chest during rapid or apprehensive breathing.
Actions
Clean the patient’s skin with alcohol or acetone.
Reposition or replace the electrodes.
Verify that the patient is comfortable, warm, and relaxed.
If wandering baseline persists, turn the baseline filter on.