Spot VS100 Vision Screener In-Service Internal Use Only
Customer Support- January 16, 2015 Page 4 of 5
Spot VS100 Vision Screener
In-Service - Quick Reference Guide
Security
4 digit pin Optional
Most people do not set this up since it is hard to communicate/maintain
Spot License Not currently used
Touchscreen calibration (Self Maintenance)
Symptoms screen needs calibration: Touch screen is not sensitive, does not
register, letters / selections do not take with one touch
Explain process below completely to user before they begin
1. Press Touch Screen
2. Press Calibrate
3. Press firmly on plus sign + that appears for just a second, then go to next plus
sign and next as they appear
4. Then plus sign with circles will appear - press those as they appear
5. System will indicate if passed or failed
6. If failed, try again
Screening Info
Instant screen: Click age on right of main screen allows immediate screening, then
put in subject data at end
Clicking Start requires entering subject data first
1. First name or Last name can be skipped.
2. You can skip gender- (but it will automatically go to male if skipped)
3. Date of birth is the only required field
4. Subject info screen- Eyewear prescription should be selected if subject is
wearing prescription, otherwise results may not be accurate
If subject wearing glasses Have subject press glasses firmly against
bridge of nose, subject should point chin downward during screening
Tips for screening
Have room environment dark as possible
Subject should not look toward light Subject should have back to ambient light
If doing screening Outdoors (under tent) - it is not recommended too much ambient
light- need black curtains
Subject should always be sitting, screener standing 3 feet away with screener
at waist height (Child may need to sit in parents lap)
Be ready to do screening in 30 seconds (default setting) or increase default setting
if needed. (Under ResultsScreening Time Out)

Other manuals for Welch Allyn Spot VS100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Spot VS100 and is the answer not in the manual?

Welch Allyn Spot VS100 Specifications

General IconGeneral
BrandWelch Allyn
ModelSpot VS100
CategoryMedical Equipment
LanguageEnglish

Related product manuals