DOC-00322-ENA.doc Page 13 of 29 ! Wescor, Inc
Nozzle Cleaning Solution, 355 mL SS-029
Nozzle Cleaning Solution, 250 mL Concentrate (dilute to 500 mL) SS-029C
Nozzle Cleaning Solution, 1.89 liter Concentrate (dilute to 3.7 L) SS-029CG
Nozzle Caps (package of six) SS-105
Preventive Maintenance Chart (pad of 24 sheets) SS-127
Disposable Cleaning Swabs for Slidepro SS-134
Thermometer for Slidepro 41-0102
Cytopro Rotor Lid Maintenance Kit SS-060
Cytopro Slide Cushions SS-109
Sample Chamber Caps for Cytopro SS-110
Fast (white) Cytopad Absorption Pads SS-111
Slow (tan) Cytopad Absorption Pads SS-112
Sample Chambers with Caps and Fast Cytopads SS-113
Sample Chambers with Caps and Slow Cytopads SS-114
Sample Chambers with Fast Cytopads SS-115
Sample Chambers with Slow Cytopads SS-116
Uncoated Custom Microscope Slides for Cytopro SS-117
Uncoated Custom Microscope Slides for Cytopro Dual Sample Chambers SS-217
POLY-L-LYSINE Coated Custom Microscope Slides for Cytopro SS-118
POLY-L-LYSINE Coated Custom Microscope Slides for Cytopro Dual Sample
Chambers
SS-218
Dual Sample Chamber Caps SS-210
Fast Cytopad Dual Sample Absorption Pads SS-211
Slow Cytopad Dual Sample Absorption Pads SS-212
Dual Sample Chambers with Caps and Fast Cytopads SS-213
Dual Sample Chambers with Caps and Slow Cytopads SS-214
Dual Sample Chambers with Fast Cytopads SS-215
Dual Sample Chambers with Slow Cytopads SS-216
d. In Vitro Diagnostic Use (8.7a and 8.4g)
The symbol on the label recognizes the diagnostic use described above (Intended purpose).
e. Storage and Handling Conditions (8.7a and 8.4h)
Reagents are marked with a storage temperature range of 18° to 31°C. Do not freeze reagents or store in direct sunlight.
Temperatures slightly outside these specified limits for a short duration will not harm the reagents.
If the Aerospray stainer will be idle for more than a week, the following procedure prevents nozzle-plugging problems when
ready to use the instrument again. Do this for each reagent line.
• Carefully loosen the ring-cap from the reagent bottle.
• Lift the dip tube out of the bottle and wipe off any remaining reagent from the tube.
• Install a cap on the reagent bottle and set aside.
• Place the dip tube in a bottle of methanol or ethanol.
• Flush at least 250 mL of methanol or ethanol through the line and spray nozzle. Leave methanol or ethanol in the line.
NOTE: If using potassium permanganate in line B, flush with 100 mL of DI water, then 150 mL of nozzle
cleaning solution (prepared SS-029C). Leave standing at least 1 hour before the methanol or ethanol flush.
• Leave flushing fluids in the reagent lines during storage. Do not run reagent lines dry.
• One at a time, remove and disassemble each spray nozzle. After removing o-ring, place metal nozzle parts in a 50 mL
centrifuge tube (provided in the Aerospray Maintenance Kit) filled with prepared SS-029C.
• Mark the tube to identify the correct reagent line. Use the provided cleaning kit tube stand to store the tubes. Place each
tube holding nozzle parts in the position corresponding with the reagent line that the nozzle is removed from. Be sure the
nozzles are returned to their original position in the stainer.
• Flush the instrument drain tube with water to prevent build-up of paper fibers, precipitates, or other materials.