50 WIKA operating instructions, model CPG1500
EN
14161966.06 03/2024 EN/DE
The Bluetooth
®
symbol on the display flashes when Wireless is switched on.
The Bluetooth
®
symbol is lit continuously when a connection has been established.
If no Bluetooth
®
connection can be established with an instrument within 30 seconds,
Bluetooth
®
is deactivated. For further information, see chapter 4.10.2 “Bluetooth®”.
6.6.2 WIKA-Cal configuration (also possible with the demo version)
1. In the WIKA-Cal, define CPG1500 as standard in “Object manager” / “Standards
(pressure)” and assign it to the workplace.
2. Open the menu item “Object manager” / “Work station”.
3. Call up the Wireless function.
⇒
The wireless monitor opens.
4. Click in the address field.
⇒
The address will be displayed automatically. If required, correct this.
The communication is working properly if the pressure value displayed on the
instrument is shown after pressing the [Test] button.
Bluetooth
®
symbol
6. Operation via menu functions