EasyManua.ls Logo

WIKA IS-3 - Page 46

WIKA IS-3
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46 WIKA operating instructions pressure transmitter, model IS-3
MM/YYYY country code based on 14095850.02 12/2014 EN/CN
EN
7.4 Finishtheadjustment(gureA)
Clampingnut(gureA)
1. Disconnect the instrument connector (3) from the power supply and the display unit.
2. Carefully insert the instrument connector (3) into the instrument, without damaging the wires or sealing. The seals
must be clean and free from damage in order to ensure the specied ingress protection.
3. Tighten the clamping nut (2).
Screw,eldcase(gureA)
Screw the screw or the case cover back in.
After the adjustment, check that the system is functioning correctly.
Recommended recalibration cycle: Annually (see chapter 8.3 “Recalibration”)
For any questions, please contact the manufacturer. See application consultant under chapter 1 “General information”
B
Field caseClamping nut Screw
S = span
Z = zero point
7. Adjusting the zero point and span

Related product manuals