GB
11258421.0901/2012GB/D
10 WIKAoperatinginstructionstemperaturetransmitterT32.xS
TheconnectionofathermocouplesensortotheT32.3S
mustbemadewithascreenedcable.Theshieldmustbe
electricallyconnectedwiththehousingofthegrounded
thermometerand,additionally,groundedtothesideof
theT32.3S.Itshouldbeensuredthatthereisequipoten-
tialbondingoninstallation,sothatnobalancingcurrent
canowviatheshield.Here,inparticular,theinstallation
regulationsforhazardousareasshouldbefollowed!
DANGER!
Danger of death caused by electric current
Uponcontactwithliveparts,thereisadirectdangerof
death.
■
Theinstrumentmayonlybeinstalledandmountedby
skilledpersonnel.
■
Operationusingadefectivepowersupplyunit(e.g.
shortcircuitfromthemainsvoltagetotheoutput
voltage)canresultinlife-threateningvoltagesatthe
instrument!
WARNING!
Onlyinstrumentsasdescribedinchapter4.2"Operation
insafety-relatedapplications"arequaliedforusein
safety-relatedapplications.Donotuseotherinstruments
insafetyorEmergencyStopdevices.
Incorrectuseoftheinstrumentcanresultininjury.
2. Safety