25WIKA operating instructions, models TRxx and TCxx
14150915.08 04/2024 EN/DE/FR/ES
5.6.4 Cleaning in place (CIP) cleaning process
Only use cleaning agents which are suitable for the seals used.
Cleaning agents must not be abrasive nor corrosively attack the materials of the
wetted parts.
Avoid thermal shocks or fast temperature changes. The temperature difference
between the cleaning agent and rinsing with clear water should be as low as
possible. Negative example: Cleaning at 80 °C [176 °F] and rinsing at 4 °C [39 °F]
with cold water.
For tank-mounted sensors the tank cleaning devices shall be positioned in such a
way that the sensor can be assessed and is perfectly cleaned.