9
them with dental lube frequently, as sterilization and repeated pulling and
inserting will reduce their working life. Change a new one once it is damaged or
worn excessively.
e) The following sterilizing methods are forbidden:
①
Put handpiece into any liquid for boiling.
②
Dip handpiece in disinfectors such as iodine, alcohol and glutaraldehyde.
③
Put handpiece into oven or microwave oven for baking.
3.2 Sterilization of scaling tips and endochuck
All the scaling tips and endochuck can be autoclaved under high temperature
and pressure.
3.3 Sterilization of torque wrench and endo wrench
3.3.1 The torque wrench and endo wrench can be sterilized under high
temperature and pressure.
3.3.2 The following sterilization ways for torque wrench are forbidden.
a) Soaked in solution.
b) Dip in iodine, alcohol or glutaraldehyde.
c) Torrefy in oven or microwave oven.
Notice: We are not responsible for any damage of the torque wrench
directly or indirectly made by any way in the above items.
3.4 Cleaning of tips, endochuck, torque wrench and endo wrench
The scaling tip, endochuck, torque wrench and endo wrench can be cleaned
by ultrasonic cleaner.
3.5 Troubleshooting and notes
3.5.1 Troubleshooting
Fault Possible cause Solutions
The scaling tip doesn't
vibrate and there is no
water owing out when
stepping on the foot switch.
The power plug is in loose
contact.
Make the plug insert to the
socket well.
The foot switch is in loose
contact.
Insert the foot switch to its
socket tightly.
The fuse in the main unit is
broken.
Contact our dealers or us.