EasyManuals Logo

Xo 4 User Manual

Xo 4
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
XO
®
4 User Guide YB-253 Version 3.10
20
9 Quality of dental air and water
Please note that the quality of dental air and water might be sub-
ject to special regulations set out in national laws.
The effective filter size on XO 4 is 5 m for air and 50 m for
water.
In order to ensure proper operational conditions for XO 4 we re-
commend the use of (an) oil-free compressor(s) equipped with
a dry air aggregate.
XO 4 fulfils the requirements in EN 1717 concerning the back-
flow of water if it is configured with XO-305 or XO-306. These
backflow protections are designed with an air gap > 20 mm and
prevent water from XO 4 from flowing back into the public supply
line. If the above named backflow protection is not mounted,
then the water supply to XO 4 must be secured against backflow
in another way.

Other manuals for Xo 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xo 4 and is the answer not in the manual?

Xo 4 Specifications

General IconGeneral
BrandXo
Model4
CategoryDental equipment
LanguageEnglish

Related product manuals