18 XO Osseo, User Guide YB-223 Version 3.00
Handpieces and contra-angels must be cleaned, maintained and sterilized strictly in accordance
with the handpiece manufacturer’s specifications and recommendations. Failure to do so may result
in reduced torque performance and shortened lifetime of the handpiece.
F.4 HANDMOTOR WITH CORD
The handmotor is supplied with a sealing plug which has an O-ring at the end. Prior to autoclaving,
push this end of the sealing plug into the open center of the handmotor shaft.
CAUTION: Failure to autoclave the handmotor with the sealing plug in place may cause
damages to the handmotor.
CAUTION: Do not lubricate through the open front end of the handmotor.
Double wrapping is recommended.
Sterilize only by steam autoclaving at a max. temperature of 135°C (275°F) and with a hold time of
at least 4 minutes.
Higher autoclave temperatures (flash autoclaving) and other sterilization methods may damage the
equipment beyond repair.
Do not leave the handmotor and cord over-night in the autoclave.
Carrier for handmotor can be autoclaved at a max. temperature of 135°C (275°F) and with a hold
time of at least 4 minutes.
F.5 SYSTEM ACCESS KEYS AND ADJUSTMENT PIN
The system access keys and the adjustment pin can be steam autoclaved together with the
handmotor and cord i.e. with the same autoclave temperature of max. 135°C (250°F) and hold time
of 4 minutes.