11
7
E:
E 0 1 2 3
Times/s 12.5 25 50 100
The higher E is, the faster speed and less stability it will be
F:
F 0 1 2 3 4
Intensity Weaker weak Middle Strong Stronger
The higher F is, the more stability and more delay it will be
8
When A of parameter H set as 0, parameter Adr,bt,tod is disabled;
[SET-2]:
Press [SET] key, it will display [set-0], press [↑] key until
it shows [set-2], then press [ENTER] key, operate step by step
refer to following table:
9
When D of parameter H set as 0, parameter A****** is disabled; If D of H is set
as 1, then indicator will output 20mA analog signal when it reaches the value
A****** set
3
[FLt EF] E-AD parameter setting
7
E- AD sampling speed
F- AD filter intensity
For example:
[FLt 22]
4
[Adr **] Address for indicator
8
(1~26)
5
[bt *] Baud rate(0~4)
0-600;1-1200;2-2400;3-48
00;4-9600
6
[tod *] Communication mode
0- Continuously send
1- Command mode
7
[Pt *]
8
[PL *] Printing language:
0- English
1- Chinese
9
[AtP *] Auto printing:
0- No auto printing
1- Auto printing
10
[A******]
2
0mA analog output weight
value set
9
11
[Prt *]
T
o print above parameter set
value or not
0- No
1- Yes
Item Display Parameter description Remark
1
[C ABCD] Control parameter
setting
10
For example:
[C 0101]

Related product manuals