5.INSTRUCTION FOR OPERATION
16
Ensure there is no load and stable indicator is on
5.1 Turning on
When indicator is powered on, it will display 0~9, then it comes
to weighing status. If the unloaded scale divates from zero
point but still within auto zero range, then indicator will
come to zero automatically;
Before Weighing
Make sure that:
a. The load cell signal cable is connected and properly
secured.
b. The scale is turned on.
5.2 Zero manually
If the unloaded scale divates from zero after auto zero when
turned on, but it’s still within manual zero range and the stable
indicator is on, press [ZERO] key, then indicator will come
to zero and zero indicator will be on;
5.3 Tare
At weighing status, when the current display is positive and
stable, press [TARE] key will tare off the current weight and
indicator will display 0, zero indicator is on
5.4 Date/time set
Press[SET] key, and then press [ENTER] key and set date, then
press [ENTER] key and set time.
5.5 Start/Stop
Item Display Parameter description Remark
1
[dC *] Decimal point(0~3) Press[ENTER]
after input
2
[E *] Division:1/2/5/10/20/
50
Press[ENTER]
after input
3
[F******] Full range Press[ENTER]
after input
4
[r 0] Keep former zero
0- Reconfirm current
zero
1- Skip zero
confirmation
nput 0,then go
to step
5(recommended)
nput 1, then go
to step 6
5 [noLoAd] Zero confirmation
16
6
[AdLoAd] Load weight
ress [ENTER]
hen stable
indicator is on
7
[000000]
nput loaded weight
value
ress [ENTER]
after input
8
[******] Display current
weight value
Exit calibration
status