You can also shut down the instrument using panel keys and soft keys.
1.
Press SYSTEM.
2.
Press the MORE soft key three times. The SYSTEM 4/4 screen is displayed.
3.
Press the SHUT DOWN soft key.
4.
Press the YES soft key. Shut down begins.
5.
After the POWER switch changes from green to orange, turn OFF the MAIN
POWER switch on the rear panel of the instrument.
Note
If for some reason the instrument fails to shut down normally, hold down the
POWER
switch
for approximately four seconds or longer to force standby mode. Note that the operating system
configuration file will not be backed up, possibly resulting in malfunctions upon start up the next
time the instrument is turned on.
Explanation
Screen when the instrument was not shut down
If the shutdown procedure was not performed after the previous session, the following
message appears after start up.
Failure to properly shut down the instrument can result in damage to the monochromator.
When turning OFF the power, always perform the shut down procedure.
Press any key to clear this message.
8 Turning the Power ON/OFF

Other manuals for YOKOGAWA AQ6370D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AQ6370D and is the answer not in the manual?

YOKOGAWA AQ6370D Specifications

General IconGeneral
BrandYOKOGAWA
ModelAQ6370D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals