<1. Introduction>
1-2
IM 01C25B01-01E
WARNING
Indicates a potentially hazardous situation which,
if not avoided, could result in death or serious
injury.
CAUTION
Indicates a potentially hazardous situation which,
if not avoided, may result in minor or moderate
injury. It may also be used to alert against unsafe
practices.
IMPORTANT
Indicates that operating the hardware or software
in this manner may damage it or lead to system
failure.
NOTE
Draws attention to information essential for
understanding the operation and features.
Direct current
1.1 Safe Use of This Product
For the safety of the operator and to protect the
instrument and the system, please be sure to follow
this manuals safety instructions when handling this
instrument. If these instructions are not heeded,
the protection provided by this instrument may be
impaired. In this case, Yokogawa cannot guarantee
that the instrument can be safely operated. Please
pay special attention to the following points:
(a) Installation
This instrument may only be installed by an
engineer or technician who has an expert
knowledge of this device. Operators are not
allowed to carry out installation unless they
meet this condition.
With high process temperatures, care must
be taken not to burn yourself by touching the
instrument or its casing.
Never loosen the process connector nuts when
the instrument is installed in a process. This can
lead to a sudden, explosive release of process
uids.
When draining condensate from the pressure
detector section, take appropriate precautions
to prevent the inhalation of harmful vapors and
thecontactoftoxicprocessuidswiththeskin
or eyes.
When removing the instrument from a
hazardousprocess,avoidcontactwiththeuid
and the interior of the meter.
All installation shall comply with local installation
requirements and the local electrical code.
(b) Wiring
The instrument must be installed by an
engineer or technician who has an expert
knowledge of this instrument. Operators are not
permitted to carry out wiring unless they meet
this condition.
Before connecting the power cables, please
conrmthatthereisnocurrentowingthrough
the cables and that the power supply to the
instrument is switched off.

Other manuals for YOKOGAWA Dpharp EJX110A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA Dpharp EJX110A and is the answer not in the manual?

YOKOGAWA Dpharp EJX110A Specifications

General IconGeneral
BrandYOKOGAWA
ModelDpharp EJX110A
CategoryTransmitter
LanguageEnglish

Related product manuals