EasyManua.ls Logo

YOKOGAWA EXAxt ZR22G

YOKOGAWA EXAxt ZR22G
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<3.SystemCongurations>
3-3
TI 11M12A01-01E
Mar. 14, 2017-00All Right Reserved Copyright © 2004, Yokogawa Electric Corporation
(2) Type: L2 (manual calibration)
EXA
ZR402G
~
<2> ZR402G Converter
Check valve
or Stop valve
<3> ZA8F Flow Setting Unit
<1> ZR22G Detector
100 to
240 V AC
Air Set
Instrument air
Calibration gas
Reference
gas
Flowmeter
Needle
valve
Zero gas cylinder
*2
Span gas
Check valve
Dust filter
Contact input
Analog output, contact output
(Digital output HART)
Signal
*1
(6-core shield cable)
Heater (2-core)
Regulator
No. Check Sketch of shadeName
1
2
3
General-purpose detector (with check valve or stop valve)
Detector with probe protector (same as above)
Detector with filter (same as above)
Note : Detector can be selected from among the following
F0302.ai
See Section 3.3 for details concerning detectors.
(3) Type: L3 (automatic calibration)
F0303.ai
EXA
ZR402G
~
<1> ZR22G Detector <2> ZR402G Converter
Signal
*1
(6-core shield cable)
Heater (2-core)
Reference gas
Calibration gas
Flowmeter
Needle
valve
Zero gas cylinder
*2
100 to 240 V AC
Check valve
<3> ZR40H Automatic
Calibration Unit
Contact input
Analog output, contact output
Digital output (HART)
lnstrument air
Regulator
Air Set
*1: Shield cable; Use shielded signal cables, and connect the shield to the FG terminal of the converter.
*2: When a zirconia oxygen analyzer is used, 100% N
2
gas cannot be used as the zero gas. Use approximately 1% of O
2
gas
(N
2
-based).

Related product manuals