<3.SystemCongurations>
3-7
TI 11M12A01-01E
Mar. 14, 2017-00All Right Reserved Copyright © 2004, Yokogawa Electric Corporation
F0308.ai
45°
[D5] [D6] [D7]
Over 700
Up to 1400
High-temperature
probe adapter
Ejector for probe adapter
(or needle valve)
High-
temperature
detector
Probe
(Note 1)
Consult us if pressure is less than –0.5 kPa
Pressure
kPa
Probe
material
Maximum
temperature
(ºC)
Conditions &
specifications
Dust level
Mounting orientation
Temperature (ºC)
Ejector for
probe adapter
Valve
Not
required
Required
Not
required
Not
required
required
Not
required
Valve Valve
Ejector for
probe adapter
Ejector for
probe adapter
–0.5 to +0.05 (Note 1)
+0.05 to +0.5 +0.5 to +5
Low (1 g/Nm
2
or less. Consult us if higher) Remarks
ZR22G-015
ZO21P-H-
ZR22G-015
ZO21P-H-
Needle valve not
supplied
ZR22G-015
ZO21P-H-
(E7046EC or E7046EN)
The high-temperature
probe adapter includes
a standard material
probe.
SUS310S
800
SUS310S
800
SiC
1400
Inconel 600
Not possible
1000
over
1000
Vertical downwardHorizontal
ZR22G-015
ZO21P-H-
E7046EC
or
E7046EN
Projection: 200 mm max.
Not standard,
please
consult us.