5-20 Parameter setting
Code Display Function Function detail X Y Z Default values
User interface
50 *RET. Auto return Auto return to measuring mode Off 0
Auto return to measuring mode On 1 1 On
51 *MODE Mode setup Manual impedance check Off 0 0 Off
Manual impedance check On 1
52 *PASS Passcode Maintenance passcode Off 0 0.0.0 Off
Note # = 0 - 9, where Maintenance passcode On #
0 = no passcode Commissioning passcode Off 0 Off
1=111, 2=333, 3=777 Commissioning passcode On #
4=888, 5=123, 6=957 Service passcode Off 0 Off
7=331, 8=546, 9=847 Service passcode On #
53 *Err.4.1 Error setting Impedance low (input 1) Soft fail 0
Impedance low (input 1) Hard fail 1 1 Hard
*Err.5.1 Impedance high (input 1) Soft fail 0
Impedance high (input 1) Hard fail 1 1 Hard
*Err.4.2 Impedance low (input 2) Soft fail 0
Impedance low (input 2) Hard fail 1 1 Hard
*Err.5.2 Impedance high (input 2) Soft fail 0
Impedance high (input 2) Hard fail 1 1 Hard
*Err.07 Temperature too high Soft fail 0
Temperature too high Hard fail 1 1 Hard
*Err.08 Temperature too low Soft fail 0
Temperature too low Hard fail 1 1 Hard
*Err.09 pH out of range Soft fail 0
pH out of range Hard fail 1 1 Hard
*Err.11 Wash recovery check Soft fail 0 0 Soft
Wash recovery check Hard fail 1
*Err.16 Call for maintenance Soft fail 0 0 Soft
Call for maintenance Hard fail 1
54 Not used
55 *CALL.M Call for maintenance Set time limit for calibration Off 0 0 Off
Set time limit for calibration On 1
Set valid maintenance period if On 250 days
56 *DISP Display resolution Set pH decimal display 0.1 pH 0
Set pH decimal display 0.01pH 1 1 0.01 pH
57-59 Not used