What to do if CYLINDER FAIL signal appears on YOKOGAWA Measuring Instruments?
- KKevin KochAug 12, 2025
If the 'CYLINDER FAIL' signal appears on your YOKOGAWA Measuring Instruments, first, ensure the PG2 pressure gauge in the control box displays the correct preset value while in cleaning mode. If the reading is low, check for loose connections and readjust accordingly. Also, remove any dirt from the cylinder barrel and apply a thin layer of grease. Degraded lubricating grease can also cause this issue, so cleaning and re-greasing are important. If the slide of the cylinder is in an upper position, refer to section 6.5.